Dosin® –a new pump device. Its construction and function

Författare: Wahlberg, A., Eriksson, U., Fuchs, L.F., Andersson, A.
Dokumenttyp: Artikel
Tillstånd: Publicerad
Tidskrift: Upsala J Med Sci
Volym: 112   83-93
År: 2007

Abstract

The cyclic flow of the Dosin® is useful in interventional radiology since it prevents clotting and increases the mixing with the blood. The flow rate varies from 0.03 mL/s to 5 mL/s and it is recommended to use fluid with a viscosity lower than 3 mPa.s. The outlet pressure is recommended below 300 mm Hg. The Dosin® must be replaced after each patient, alternatively after maximum 24h continuous use.